Abstract
BackgroundHigh placement of a cementless acetabular cup is often required to have sufficient bone coverage in hip dysplasia. We examined whether a high hip center decreases the postoperative range of motion (ROM) in total hip arthroplasty. MethodsUsing a computer software, ROM and bone coverage were examined in 32 patients with unilateral osteoarthritis of the hip with Crowe type II or III hip dysplasia. The cup was placed at the anatomic hip center and moved vertically in 5-mm increments. Vertical center of rotation (V-COR) was defined as the distance from the head center to the interteardrop line. The required ROM was defined as flexion ≥110°, internal rotation (IR) at 90° flexion ≥30°, extension ≥30°, and external rotation ≥30°. We determined the V-COR to satisfy the required ROM and cup center-edge angle (Cup-CE) ≥0°. ResultsOnly 40.6% of the patients satisfied Cup-CE ≥0° at the anatomic hip center. Bone coverage increased with the peak at 30, 35 mm of V-COR (90.6% satisfied Cup-CE ≥0°) and decreased at ≥40 mm. Higher V-COR resulted in lower range of flexion and IR, but increased the range of extension and external rotation. The cutoff value from the receiver-operating characteristic curve for both flexion and IR was a V-COR of 35 mm. ConclusionThe higher hip center gained more bone coverage but decreased the range of hip flexion and IR. Acceptable V-COR in hip dysplasia was around 35 mm, about 20 mm above the true hip center, in terms of ROM.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.