Abstract

Micro electro-mechanical and NANO technologies are sensibly reducing circuit and geometrical errors in accelerometer sensors. These sensors are often used as portable inclinometer sensors for trunk flexion monitoring in clinical applications. In this case, the rotating trunk generates centripetal acceleration, an error source that technological efforts cannot eliminate. This study analysed the effect of this source for typical human monitoring conditions by simulations and clinical validation using a wearable device with rate gyroscopes and accelerometers (Giansanti and Maccioni 2005 Physiol. Meas. 26 689–705). Results showed that this error source did not affect long-term monitoring applications (Mathie et al 2004 Physiol. Meas. 25 R1–R20) but in the short-term monitoring caused a mean angular error equal to 0.96° for the most critical single-task application represented by the sit-to-stand.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.