Ultrasound bladder vibrometry (UBV) parameters have been shown in previous studies to strongly correlate with measurements from urodynamic studies. Just like urodynamic studies, UBV can be performed in supine and sitting positions. The objective of this study is to compare UBV parameters obtained in the two different positions using statistical methods. We recruited eight volunteers with healthy bladders for this purpose. The elasticity, group velocity squared and thickness of the bladder were the UBV parameters of interest, and their values were recorded at different bladder volumes for each volunteer. The results presented indicate that the measurements made in the two positions are in agreement using the Bland–Altman method and a parameter q which compares the values at each bladder volume for each volunteer. UBV parameters were also repeatable for measurements recorded in the supine and sitting positions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.