Abstract

Background and purpose The underdose near air cavities in the head and neck region at photon energies of 4 MV and 6 MV was studied in search for clinical advantages of the 4 MV over 6 MV treatments. Materials and methods The on-axis and off-axis dose distributions were measured with a parallel-plate ionization chamber and films in polystyrene phantoms containing an air cavity of appropriate size based on the results of computed tomography scans. Results Although most results are similar for both energies, the 4 MV photon beams give a somewhat smaller underdose effect and a faster re-build up than the 6 MV. For both energies a significant underdose effect was observed at the edge of the field in the larynx phantom. This proved to be true for small and large fields, for smaller and larger cavities, for one-beam as well as parallel-opposed beams. Conclusion For most clinically relevant situations there is no remarkable benefit in the use of either of the two energies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.