Abstract

In this issue of Journal of Diabetes Science and Technology, Baumstark et al. evaluated the analytical performance of a bench-top laboratory glucose analyzer (SUPER-GL) intended for replacement for the YSI2300-STAT analyzer, that served for several decades as a comparator method in clinical and analytical studies of blood glucose monitoring systems (BGMS). The authors concluded that the SUPER-GL's overall performance is comparable to that of YSI2300-STAT, and has the potential to be a candidate comparator analyzer. However, the question is if we need to recommend as a "comparator method," a specific device, that measure glucose using the same analytical method with most BGMS. In this analysis we present our point of view hoping to generate a discussion on the necessity for such a replacement.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call