Abstract
Aerosol mitigation equipment implemented due to COVID-19has increased noise levels in the operating room (OR) during otolaryngological procedures. Intraoperative sound levels may potentially place personnel at risk for occupational hearing loss. This study hypothesized that cumulative intraoperative noise exposures with aerosol mitigation equipment exceed recommended occupational noise exposure levels. Sound levels generated by the surgical smoke evacuator (SSE) during adenotonsillectomy were measured using a sound level meter and compared to surgery without SSE. Thirteen adenotonsillectomy surgeries were recorded. Mean sound levels with the SSE were greater than the control (72 ± 3 A-weighted decibels (dBA) vs. 68 ± 2 dBA; p=0.015). Maximum noise levels during surgery with SSE reached 82 ± 3 dBA. Surgeons performing adenotonsillectomy with aerosol mitigation equipment are exposed to significant noise levels. Intraoperative sound levels exceeded international standards for work requiring concentration. Innovation is needed to reduce cumulative OR noise exposures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.