Abstract
Our objective was to assess the effect of orthopedic metallic prostheses on radiation dose associated with MDCT using z-axis automatic tube-current modulation and a fixed tube current in a phantom and patient study. Z-modulation is associated with a 34.1% increase in the mean tube current-time product and no change in the extent of streak artifacts in patients with a metallic prosthesis, compared with patients without a prosthesis. However, compared with the fixed-tube-current technique, z-modulation is associated with a 28.9% decrease in the mean tube current-time product.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.