Abstract

The use of dynamic external fixators for the treatment of long bone fracture is widespread and well accepted. It is claimed that dynamization, i.e. small micromovements of compression/distraction at the fracture site, can be produced by allowing sliding of an inner rod within an outer housing. However, as the forces on the fixator are not direct but transferred from the bone via bone pins, there is a bending moment on the fixator. This produces “self-locking” and effectively prevents axial movements. We have studied the effect of this moment on the binding properties of the Orthofix system. The amount of movement at a simulated fracture site allowed before this locking occurs was measured and its implications discussed. It would appear that true axial dynamization does not take place using this system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.