Abstract
The spatial distribution of the activity in a flame-sealed PTB ampoule containing 226Ra solution, incompletely filling the ampoule, was studied by autoradiography and by taking well-collimated gamma-ray spectra in 23 sections of the ampoule. A complex activity profile was observed and used for the Monte Carlo simulation of the ampoule by the GESPECOR code. The computed values of the efficiency transfer factors are in agreement with the experimental data. As a byproduct, the radon partition ratio in the solution–air phases was obtained.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.