Abstract

To assess the validity of using the calibration ruler for correcting magnification of linear measurements and to explore and compare the vertical and horizontal magnification of four digital cephalometric units. An acrylic box was imaged at seven sagittal positions using four digital cephalometric units: Orthopantomograph OC100, Orthopantomograph OC200, Sirona Orthophos CD, and Sirona Orthophos DS. The true linear lengths of the phantom, corrected, and uncorrected linear lengths on the images were measured and compared. The validity of measurements using the calibration ruler was assessed. The magnification values and distortion indices were calculated and compared among the four cephalometric units. For linear measurements on the mid-sagittal plane and averaged linear measurements on bilateral symmetric sagittal planes, the bias 1.96 STD of the calibration ruler ranged from 1% to 2% for the four cephalometric testing units. For linear measurements on the single lateral sagittal plane, the bias 1.96 STD ranged from 3% to 6%. The vertical scanning charge-coupled device cephalometric unit produced the greatest distortion, ranging from 1.029 to 0.964. The metal millimeter calibration ruler is an accurate reference for linear measurement magnification correction. Because of unpredictability and machine specificity, the magnification and distortion of a cephalometric unit should be calibrated for the estimation of cephalometric measurement error.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.