Abstract
To explore high-field EPR in biological applications we have compared measurements of dynamics with X-band (9 GHz) and W-band (94 GHz) saturation transfer EPR (ST-EPR) and distance determination by X and W-band DEER. A fourfold increase of sensitivity was observed for W-band ST-EPR compared with X-band. The distance measurements at both fields showed very good agreement in both the average distances and in the distance distributions. Multifrequency EPR thus provides an additional experimental dimension to facilitate extraction of distance populations. However, the expected orientational selectivity of W-band DEER to determine the relative orientation of spins has not been realized, most likely because of the large orientational disorder of spin labels on the protein surface.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.