Abstract
The purpose of this study was to assess the effectiveness of a novel radiation-independent aiming device for distal locking of intramedullary nails in a human cadaver model. A new targeting system was used in 25 intact human cadaver femora for the distal locking procedure after insertion of an intramedullary nail. The number of successful screw placements and the time needed for this locking procedure were recorded. The accuracy of the aiming process was evaluated by computed tomography. The duration of the distal locking process was 8.0 ± 1.8 minutes (mean ± SD; range, 4-11 minutes). None of the screw placements required fluoroscopic guidance. Computed tomography revealed high accuracy of the locking process. The incidence angle (α) of the locking screws through the distal locking holes of the nail was 86.8° ± 5.0° (mean ± SD; range, 80°-96°). Targeting failed in 1 static locking screw because of a material defect in the drilling sleeve. This cadaver study indicated that an aiming arm-based targeting device is highly reliable and accurate. The promising results suggest that it will help to decrease radiation exposure compared with the traditional "free-hand technique."
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.