Abstract

This laboratory-based study compared distal fibula simple oblique fracture fixation with one-third tubular plate with and without a single lag screw to determine which was mechanically more stable. A control group fixed with a limited contact dynamic compression plate was also tested. Biomechanical testing of 30 osteotomised saw bones under lateral bending and torsional forces was performed. There was no significant difference between the mean lateral bending and mean torsional stiffness between the fixation with tubular plate and lag screw and tubular plate alone. Limited contact dynamic compression plate conferred the best stability in lateral bending and torsion, as expected.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.