Abstract

The left ventricular assist device Impella 5.0 (Abiomed Inc, Danvers, MA) has become widely accepted as a temporary mechanical circulatory support for patients in cardiogenic shock. The Impella 5.0 is placed through an anastomosed graft. When removing the device, blood clot formation has been noted in the anastomosed graft. The blood clot has been reported to dislodge and embolize distally, causing acute limb ischemia. Here, we present our simple, inexpensive, and effective "loop and snare" technique for safer device removal, preventing distal embolic complications. In our experience of 6 patients who had Impella 5.0 removal with this technique, the distal embolic complication was 0%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.