Abstract

The analysis of some Italian mineral waters by ICP-MS has revealed errors in the determination of As and Cr in natural effervescent or carbonated waters due to the presence of dissolved inorganic carbon (DIC). This leads to overestimate As and Cr in 1% (v/v) HNO 3 acidified samples, analysed within 1–2 h after the acidification. The overestimation of As concentration is caused by matrix interferences producing a signal enhancement due to the presence of dissolved inorganic carbon. This effect is analogous to that observed in the presence of organic carbon and occurs at millimolar DIC levels. The overestimation of Cr concentration is due to the 40Ar 12C + species interfering with 52Cr + despite the use of the octopole reaction system. The optimization of the He flow in the collision cell can solve the latter problem, but the required increase in the flow rate decreases the sensitivity of the ICP-MS technique. The observed effects in CO 2 rich mineral waters and artificial NaHCO 3 solutions suggest that 5–10 mM DIC levels may affect the determination of As and Cr concentration in thermal waters, rivers, lakes and groundwaters.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.