Abstract

Dissociation of a modular tibial insert from the metal baseplate is a rare complication after revision total knee arthroplasty. We herein report an unusual presentation of a polyethylene insert failure that occurred in a posterior-stabilized constrained total knee arthroplasty. The polyethylene insert was posteriorly displaced, which has never been reported in the literature. The assumed mechanism is disengagement of the locking screw after repetitive motion. During knee full extension while standing from a chair, a sudden downward force on the anterior half of the insert caused posterior lift-off of the insert from the tibial baseplate and promoted a sliding force to push the insert posterior when the locking mechanism failed. This unique complication in a fully constrained posterior-stabilized implant should be considered as a cause for implant failure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.