The large variety of high-precision unique blackbody sources: those operating at fixed temperatures provided by phase transitions of metals and metal-carbon eutectics, and variable-temperature ones had been designed in VNIIOFI for high-precision radiometry, radiation thermometry and spaceborne remote sensing within a 100 to 3500K temperature range. Paper reviews the blackbodies (BBs) ranged to low, middle and high temperatures, and describes spectral radiance and irradiance calibration facilities on the base of these BBs in IR and V-UV spectral ranges. The latest investigations of high-temperature fix-points based on metal-carbon eutectics Re-C (2748K) demonstrated an excellent reproducibility of freezing plateau (up to 0.01% in terms of radiation temperature) between series of measurements/crucibles, and about 0.003% within a sample measurement session, i.e. better than 100mK. Further Re-C (spectral irradiance measurements) and TiC-C (3057° C) eutectics are being investigated for use as high-stable radiance/irradiance sources above the conventionally assigned values of temperatures of ITS-90.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.