Abstract
We have designed, fabricated and tested a real-time micro polymerase chain reaction (microPCR) system. It consists of a microscope glass cover slip placed on top of a micromachined silicon chip integrated with a heater and a temperature sensor. A single microL of a sample containing DNA was placed on the glass and encapsulated with mineral oil to prevent the evaporation of water, thus forming a virtual reaction chamber (VRC). The PCR chip required half a second to heat up from 72 to 94 degrees C and two seconds to cool from 94 to 55 degrees C, corresponding to a cooling rate of -20 K s(-1). The real-time PCR yield was determined by a fluorescence method. The melting curve analysis method as well as capillary electrophoresis was performed to determine the purity of the PCR product. As the glass slip is disposable, cross-contamination from sample to sample is eliminated. The total cost of running the PCR is given by the value of the cover slip and its treatment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.