Abstract

A disposable microfluidic device was constructed by conveniently integrating one poly(methyl methacrylate) board with four reservoirs and one fractured fused-silica capillary with 50 microm i.d. and 7.5 cm total length on a printed circuit board for applying sampling and separation voltages. The disposable microfluidic device combined with a home-made ultraviolet workstation could be conveniently used for efficient screening and quantitative detection of microg mL(-1) illicit drugs. Using eight illicit drugs as models, they could be baseline-separated within 240 s with the separation efficiency up to 600,047 plates m(-1) at the designed device. The novel device and proposed protocol were successfully used to screen illicit drugs in human urine. This work presented a simple and low-cost method to fabricate the microfluidic device and provided a powerful way for sensitive and specific multi-screening of different drugs with high resolution, fast separation and low-cost.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.