Abstract
This study reports the preparation of disposable microcolumns with welded metal frits for the first time. First, the bottom of glass-lined stainless-steel tubing of 30 cm length, 1.6 mm od, and 0.5 mm id was welded with a stainless-steel screen frit of 1.6 mm diameter. A micro-welding machine was used for this. Next, the column was connected to a slurry packer and packed with porous silica particles. Then, the inlet of the column was carefully welded with another frit. The column was tested for separation of a test mix composed of phenol, 2-nitrophenol, acetophenone, aceanilide, and benzamide. Another column of the same physical dimension was also prepared with frits that were not welded to the column. The chromatographic performances of the two groups of columns (welded frits versus non-welded frits) were examined. The columns of welded frits showed ca. 18% better separation efficiency (number of theoretical plates) than those of non-welded frits.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.