Abstract

Disposable electrochemical immunosensor for the detection of carcinoembryonic antigen (CEA) in saliva and serum was developed. Monoclonal anti-CEA antibodies (αCEA) were covalently immobilized on polyethyleneimine wrapped multiwalled carbon nanotubes screen-printed electrode. A sandwich immunoassay was performed with CEA and αCEA tagged ferrocene carboxylic acid encapsulated liposomes (αCEA-FCL). The square wave voltammetry (SWV) was employed to analyze faradic redox responses of the released ferrocene carboxylic acid from the immunoconjugated liposomes on the electrode surface. The magnitude of the SWV peak current was directly related to the concentration of CEA. The calibration curve for CEA concentration was in the range of 5×10−12 to 5×10−7gmL−1 with a detection limit of 1×10−12gmL−1 (S/N=3). This method provides a high precise and sensitive determination of CEA in human blood serum and saliva samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.