Abstract
We evaluated the effect of orthopedic-metal artifact reduction (O-MAR) for metal artifact in computed tomography with 73 simulated seeds for brachytherapy in different sizes of display field of view (DFOV) obtained by helical scan under the same clinical scan condition. The metal artifacts were analyzed with the Gumbel's method by changing DFOV sizes 80 mm, 160 mm, and 320 mm. Gumbel distribution, scale parameter (γ), and location parameter (β) of the metal artifacts with O-MAR were compared with that of the metal artifacts with filtered back projection (FBP). In conclusion, it was considered that the effect of metal artifact reduction with O-MAR was influenced by DFOV size in this study. The reduction rates of scale parameter (γ) were 22.3%, 21.3%, and 10.0% in DFOV 80 mm, 160 mm, and 320 mm, respectively. The reduction rates of location parameter (β) were 27.4%, 23.4 %, and 9.8%. Therefore, the effect of metal artifact reduction with O-MAR showed the tendency of increasing with decreasing DFOV size.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.