Abstract
In this study a displacement current capacitance sensor (DCCS) for scanning capacitance microscopy (SCM) is introduced. It can be used for both intermittent contact (IC) and contact-SCM operation. Based on I/ V conversion and subsequent lock-in amplification a displacement current can be detected and used as a measure for dopant concentration. Therefore a periodic variation of the AFM tip substrate capacitance is required. This can be achieved either by a periodic tip oscillation (IC-SCM) or an applied AC voltage between tip and sample (contact-SCM). The advantage of the DCCS is the linearity, which makes it possible to detect absolute dopant concentrations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.