Abstract

This article describes a new procedure for multi-element preconcentration of heavy metal ions, specifically of Cr(III), Co(II), Ni(II), Cu(II), Zn(II) and Pb(II) ions. The method is based on dispersive micro solid-phase extraction (DMSPE) of the metal complexes of 2-(5-bromo-2-pyridylazo)-5-diethylaminophenol (5-Br-PADAP) by graphene oxide nanoparticles prior to energy dispersive X-ray fluorescence spectrometric (EDXRF) determination. The effects of pH, amount of graphene oxide, concentration of complexing reagent, sample volume and sorption time were optimized. The influence of commonly encountered other ions was also investigated. Under optimal conditions, the calibration plots cover the 2 to 150 ng mL−1 range for each element. The precision (at a 20 ng mL−1 level for n = 10) is lower than 4.8 %, and the detection limits range is from 0.07 to 0.25 ng mL−1. The DMSPE-EDXRF procedure was successfully applied to the determination of heavy metal ions in water and spiked water samples with recoveries between 94.4 and 103.5 %.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.