Abstract
The magnetic profile of the wiggler to be used in the planned Weizmann Institute FEL has been measured using the pulsed-wire method. The main transverse deflection pattern caused by an electrical current pulse in a wire placed along the wiggler was sometimes accompanied by minor faster and slower parasitic components. These components interfered with the main profile, resulting in distorted mapping of the wiggler magnetic field. Their periodical structure being very close to the main pattern could not be easily resolved by applying a numerical Fourier transform. A strong correlation between the wire tension and the amplitude of the parasitic patterns was found. Significant damping of these oscillations was achieved by applying high enough tension to the wire (close the yield point), allowing to disregard their contribution to the measurement accuracy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.