Abstract

Coronary stent loss is a rare but serious complication during interventional cardiology. This complication occurs not only in the intracoronary area but also in the extracoronary area, such as the aortic root or the left ventricle. An 83-year-old man with stable angina had a stent inserted into a heavy calcific left anterior descending artery. The stent was lost twice during the procedure. The first stent was dislodged from the left main coronary artery to the proximal left anterior descending artery, and the second stent migrated to the aortic root following separation from the balloon. We successfully redeployed the first stent at the dislodged site and retrieved the second stent using a goose-neck snare after moving the stent to the descending aorta. These steps circumvented the need for the patient to undergo emergency cardiovascular surgery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.