Abstract

Dislocation after bipolar hemiarthroplasty is rare. In classical dislocations of bipolar prosthesis, the prosthetic femoral head escapes from the natural acetabular cavity. Two cases of a bipolar hip prosthesis are presented in which the dislocation occurred between the cup and the metallic ball head 4 and 12 months postoperatively. The reason for the dislocation was fracturing of the edge of the polyethylene ring - either due to impingement at the full flexion position of the hip or to dislocating forces in extreme positions of the metallic ball head. To avoid this complication we recommend installation of the notch of the polyethylene ring be at the peripheral half of the acetabular cup.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.