Abstract

The procedure for determining the 67Ga disintegration rate by a primary method is described. The proposed triple 4πβ–γ coincidence system consists of a thin window gas-flow 4π proportional counter (PC) coupled to a NaI(Tl) scintillator and a HPGe crystal. Independent pulse height and occurrence time information is provided for the three detector outputs by means of a Software Coincidence System. Separate spectrometry measurements with a n-type reversed electrode coaxial Ge detector (REGe) were performed for obtaining gamma-ray emission probabilities per decay. Accurate values of disintegration rate, gamma-ray emission probabilities and the metastable half-life were achieved.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.