Abstract

Objective To investigate the effects of two different methods for the disinfection of ventilator circuits.Methods The ventilator circuits used by patients were disinfected by the soaking disinfection with 5% available chlorine disinfectant (experimental group)and the strict program of the full automatic cleaning- disinfecting machine (control group),respectively.Samples from circuits were cultured after disinfection,and the numbers of bacterium colony forming unit (CFU) were counted.Results Comparing the exeperimental group and control group,the difference was statistically significant ( P <0.01 ).Conclusions The sterilizing effects of the ventilator circuits with the full automatic cleaning and disinfecting machine are more effective and safer.The machine can prevent the cross infection and standardize clinical work. Key words: Ventilator circuit; Full automatic cleaning and disinfecting machine; Cleaning; Disinfection

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.