Regenerated, ozone-disinfected dialyzers were use for a total of 190 hemodialyses in 35 patients without the occurrence of any complication or pyrogen reactions; the efficiency of hemodialysis was identical with that of new or formalin-disinfected dialyzers. The priming fluid contained in both compartments of ozone-disinfected dialyzers proved to be sterile. The ozone did not damage the dialyzer membrane or the supporting structure of dialyzer. Ozone can effectively substitute formalin and other chemical sterilants as a disinfectant of regenerated capillary dialyzers. The process can also be automated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.