Abstract
PurposeThe aim of this study was to evaluate the flow volume and resistive index parameters of the brachial artery in the functioning and dysfunctioning radiocephalic and brachiocephalic hemodialysis arteriovenous fistula (AVF).Methods82 patients were distributed into three groups according to their hemodialysis function - as normal function, decreased pump flow and increased venous pressure. Flow volumes and resistive indexes (RI) of the brachial artery of radiocephalic and brachiocephalic AVFs were measured by Doppler ultrasound. Flow parameters of the groups were compared.ResultsA statistically significant difference was found between the normal and decreased pump flow groups in terms of flow volume and resistive index values. 770 ml/min flow volume of the brachial artery has a 94% sensitivity and 84% specificity and 0.52 value of RI has an 89% sensitivity and 88% specificity in the the differentiation of normal and decreased pump flow groups.ConclusionDoppler parameters of the brachial artery such as flow volume and RI can provide valuable information about AVF function.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.