Abstract

ABSTRACTScanning Electron Microscope–Mineral Liberation Analysis (SEM–MLA) can be used to discriminate between hematite and magnetite in iron ores. However, achieving backscattered electron (BSE) segmentation between the two minerals is difficult for particles ≤75 µm using typical preparation and analysis methods for the MLA method based on a tungsten filament SEM (Quanta 400) with 25 kV high voltage. Preparing iron ore sample mounts using a slow-speed polishing method, and conducting the experiment on a field emission gun SEM–MLA (Quanta 650) with the high voltage setting lowered to 15 kV reduces instrument noise and results in very clean BSE images and segmentation. This method requires new X-ray standards for each mineral at 15 kV because of major changes in the spectra at lower kV. However, once these X-ray spectra are added to the mineral reference list, effective segmentation can be achieved and an accurate analysis obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.