Abstract
It is shown that for discrete liquid-helium level-meter sensors, based on TVO resistors with a rated power of 0.125 W and rated resistance value of 1 kΩ, one can attain a helium-level measurement error of ±0.5 mm for the horizontal TVO resistor disposition and ±2 mm for the vertical one.