Abstract

It is shown that for discrete liquid-helium level-meter sensors, based on TVO resistors with a rated power of 0.125 W and rated resistance value of 1 kΩ, one can attain a helium-level measurement error of ±0.5 mm for the horizontal TVO resistor disposition and ±2 mm for the vertical one.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call