Abstract This paper deals with a problem in optimizing the flow rate of the mobile phase in countercurrent chromatography (CCC), A set of data including theoretical plate number (N), resolution factor (Rs), % retention, etc. are obtained from the separation of three indole auxins in a two-phase solvent system of n-hexane / ethyl acetate / methanol / water (1:1:1:1) by applying the flow rates of the mobile phase at 0.5- 8.0ml/min. The experimental results revealed a disagreement between the efficiencies expressed in parameters N and Rs at the high flow rates of the mobile phase in the CCC separation. Namely, the partition efficiency expressed in terms of Rs decreases with an increased flow rate of the mobile phase, while the N rises with increasing the flow rate. Since Rs represents the actual separation of the solute peaks, the results indicate that the N used in high performance liquid chromatography is not always a reliable measure in CCC separation. However, the application of a high flow rate yields b...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.