Abstract
The techniques of fluorescent antibody (FA) and counterimmunoelectrophoresis (CIE) were compared with the Neufeld test (quellung reaction) for typing of Streptococcus pneumoniae. A total of 88 isolates were examined by these three methods. Pool-, type-, or group-specific pneumococcal antisera were used in all three methods. Each isolate was initially tested with polyvalent antisera and all of the nine pools of antisera. Selection of the type- or group-specific antisera depended upon the reaction of the isolate with the pool sera. Sixty-eight of 88 (77.3%) isolates were positive using pool or typing sera and were correctly typed by CIE, while FA was found to be accurate for only 61 of 88 (69.3%) isolates. Positive reactions with more than one pool- or type-specific antisera, or no reaction, were seen with several of the isolates with both techniques. Even though CIE and FA are rapid and simple techniques, microbiologists should be cautious when utilizing them for typing of S. pneumoniae because of the discrepancies observed in this study.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.