Abstract
We report an extremely rare case of inner head displacement from the stem of a bipolar hip prosthesis (BHP). An 88-year-old woman underwent BHP implantation for right femoral neck fracture. However, severe right hip joint pain occurred 12 days after surgery. A plain radiogram film revealed displacement of the inner head from the neck of the stem, accompanied by sinking of the stem. At reoperation, the inner head was disassembled from the stem, and Vancouver type A1 fracture was confirmed. Disassembly may have been caused by the pumping phenomenon or micromovement of the stem due to periprosthetic fracture. To our knowledge, this is the first report about disassembly of the BHP inner head, probably due to periprosthetic fracture.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.