Abstract

ObjectiveLead orientation is a new degree of freedom with directional deep brain stimulation (DBS) leads. We investigated how prevalent deviations from the intended implantation direction are in a large patient cohort. MethodsThe Directional Orientation Detection (DiODe) algorithm to determine lead orientation from postoperative CT scans was implemented into the open-source Lead-DBS toolbox. Lead orientation was analyzed in 100 consecutive patients (198 leads). Different anatomical targets and intraoperative setups were compared. ResultsDeviations of up to 90° from the intended implantation direction were observed. Deviations of more than 30° were seen in 42% of the leads and deviations of more than 60° in about 11% of the leads. Deviations were independent from the neuroanatomical target and the stereotactic frame but increased depending on which microdrive was used. DiscussionOur results indicate that large deviations from the intended implantation direction are a common phenomenon in directional leads. Postoperative determination of lead orientation is thus mandatory for investigating directional DBS.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.