Abstract
A spark-based, solid-sampling cell is described for inductively coupled plasma mass spectrometry (ICP-MS). The cell is devised for the direct sampling of gold and silver beads produced by the classical lead fire assay procedure. The sampler produces a solid aerosol composed of submicron-sized vapor condensates and small (< 2 μm) spherules. In contrast to solution nebulization, the mass spectrum for spark-ICP-MS is relatively free of interfering metal oxide, polyatomic, and multiply-charged ions. The measurement precision is 3% RSD for Pt, Pd, and Rh preconcentrated into fire assay beads, but is 6% RSD for Ir due to its heterogeneous distribution in a silver bead. Detection limits determined for Pt, Pd, Rh, and Ir in fire assay beads range from 0.6 μg/g (Pt) to 1.2 μg/g (Pd). Calibration curves for these elements are linear up to the highest concentration in the bead studied (2000 μg/g). The quantitative potential of the method is evaluated using the South African Reference Material (SARM-7) geologic standard.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.