Abstract

Methods for the direct solid sample trace element determination in complex composed inorganic powder samples, e.g. sediments, soils, rocks, minerals and similar technical materials, are systematically investigated by SS-ETV-ICP-AES and SS-GF-AAS. High-grade homogeneity of the sample powder even at milligram and sub-milligram levels is required due to small sample weights. All the investigations are carried out with CRMs and RMs. SS-ETV-ICP-AES, which has been improved by high temperature vapor halogenation, is used for the determination of Cr, Cu, Ni, Pb, and Zn in the trace element range 500 μg/g. No significant influence of matrix components is observed using this method. SS-GF-AAS is used for the determination of Cd, Pb, and Zn in the 0.006 to 150 μg/g range. Precision and accuracy of SS-GF-AAS are improved by application of 3D-calibration plots in comparison to conventional 2D-calibration. In this way, the two methods SS-ETV-ICP-AES and SS-GF-AAS complement one another concerning their analytical performance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.