Abstract
Ramosetron is an enantiopure active pharmaceutical ingredient marketed in Japan since 1996 and later in a few Southeast Asian countries predominantly as an antiemetic for patients receiving chemotherapy. In this study, a simple and rapid high-performance liquid chromoatography method for the separation of ramosetron and its related enantiomeric impurity by using hydrophilic interaction liquid chromatography mode is presented. Chiral resolution was performed on an analytical column (100mm × 4.6mm id) packed with 3 μm particles of cellulose-based Chiralpak IC-3 chiral stationary phase. Using a mobile phase containing acetonitrile-water-diethylamine (100:10:0.1, v/v/v) and setting the column temperature at 35°C, the resolution value was 7.35. At a flow rate of 1mL/min, the enantioseparation was completed within 5min. The proposed method was partially validated and it has proven to be sensitive with limit of detection and limit of quantitation of the (S)-enantiomer impurity of 44.5 and 133.6ng/mL.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.