Abstract

A direct sample fraction deposition method was developed for off-line size-exclusion chromatography (SEC)/matrix-assisted laser desorption/ionization (MALDI) time-of-flight mass spectrometry. By using electrospray, the SEC eluent, together with a suitable matrix solution added coaxially, was directly deposited on the MALDI plate. Owing to the formation of very small droplets in electrospray, solvent evaporation is much faster. The fractionation volume in narrow-bore SEC, which can directly be collected in one MALDI spot, can easily be optimized in the range of a few microlitres. In addition, fairly homogeneous sample spots were obtained. The possible influence of composition variation of the SEC effluent on the analytical results using direct fraction deposition was investigated; no substantial effects were observed. The applicability of the method was demonstrated by characterizing a broad poly(methyl methacrylate) sample. Copyright 2000 John Wiley & Sons, Ltd.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.