Abstract
Abstract A potentiometric analyzer, developed to directly measure potassium and sodium concentrations, was evaluated on aqueous solutions, whole blood, serum, and plasma. The analyzer consists of three electrodes—sodium, potassium, and reference—mounted in a flow-cell configuration, and a modified digital pH meter, which displays the ion concentrations directly. Volume of the cell is about 100 µl. The electrodes equilibrate to each new sample in less than a minute. Calibrated with aqueous standard solution, the system yields essentially the same values for whole blood and the corresponding plasma, and results agree with those obtained for the same samples of plasma by flame photometry.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.