Abstract

A double-focusing sector field ICP MS was optimized for the direct simultaneous determination of Ag, Al, Ba, Ca, Cd, Co, Cr, Cu, Fe, Mg, Mn, Mo, Ni, Pb, Sn, Ti, V in organic solutions. Polyatomic interferences originating from the carbon-rich matrix were completely eliminated at a resolution of 4000 allowing the detection limits at the low pg g−1 level to be obtained (typically one order of magnitude lower than using a quadrupole ICP MS). A method for the routine comprehensive trace element analysis of xylene solutions of oil samples using external calibration was developed. It was validated by the analysis of three certified reference materials and applied to the analysis of 2 gas condensate samples and 4 oil samples of different origins.