Abstract
The fundamental equation for evaluating the hysteresis losses P h = ??? H dB, transformed for the use of electrical units and approximated as numeric integral Ph = W 1 /W 2 2/1 m A ???? -J ??+J J is sufficiently correct ???? is constant and very small. By continuously changing the magnetizing current in the primary winding w 1 of Epstein frames or ring cores (electronically controlled so that the induced voltage in the secondary winding W 2 does not exceed a fixed limit) the flux ?? can be determined by means of a voltage-to-frequency-converter that generates a pulse train whose frequency is proportional to the input voltage and summing the pulses with an electronic counter. These pulses are scaled to groups of constant number (representing ???? = constant) by means of a preset counter that generates trigger pulses giving commands to a sample and hold digital voltmeter in order to measure the instantaneous values of the current as voltage drop across a standard resistor in the primary circuit. An electronic adder with Nixie display totalizes the current values according to the sign. The measuring procedure takes about 40 s, the accuracy is in the order of some parts in 1000.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.