Abstract

We describe a new device in which an accurately measured force is applied to individual adherent cells while the topography of the adhesion zone is simultaneously monitored. The force is applied via a flexible glass micropipette, attached by suction to the cell under study, and is calculated directly from the measured pipette deflection. Regions of close contact in the adhesion zone are observed using interference reflection microscopy. We have used the device to measure the force required to detach human red blood cells from hydrophobic and hydrophilic glass surfaces, and to detach Dictyostelium discoideum amoebae from a hydrophobic glass surface. The measured forces per unit length of contact perimeter are within an order of magnitude of the tensions required for membrane rupture.