Abstract

The direct quantitative measurement of the concentration of water-dissolved saturated hydrocarbons using a portable membrane-inlet precalibrated magnetic mass-spectrometer is suggested. The particularity of the instrument’s response to the presence of desired compounds in a test sample is taken into consideration. The feasibility of using this portable instrument in real-time oil and gas search on the continental shelf is discussed. Laboratory test data for the methane, ethane, propane, and butane concentrations in air and water samples are presented, from which the sensitivity of the instruments to the components, are determined. The gas concentrations in the water samples lie in the range from several parts per billion (ppb) to several hundred ppb. These values are close to the lower limits of prospecting criteria as applied to oil fields.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.