Abstract

The aim of this prospective study was to evaluate the Nobel Direct and Nobel Perfect one-piece implants (OPIs) when used for immediate function. Forty-eight patients were provided with 115 OPIs for loading with a provisional crown or a bridge within 24 h and followed for at least 12 months with clinical and radiographic examinations. A group of 97 patients previously treated under identical conditions by the same team with 380 two-piece implants (TPIs) for immediate loading in the mandible and maxilla served as the reference group. Six (5.2%) OPIs failed during the follow-up due to extensive bone loss. Five (1.3%) implants failed in the reference group. After 1 year, the mean marginal bone loss was 2.1 mm (SD 1.3) for OPIs and 0.8 mm (SD 1) for TPIs. 20% of OPIs showed more than 3 mm of bone loss compared with 0.6% for TPIs. When compensating for vertical placement depth, OPIs still showed a lower marginal bone level and thus more exposed threads than TPIs. Depending on the criteria used, the success rate for OPIs was 46.1% or 72.2% compared with 85% or 91.6% for TPIs. The Nobel Direct and Nobel Perfect OPIs show lower success rates and more bone resorption than TPIs after 1 year in function. Factors such as implant design, insertion depth, rough surface towards the mucosa, in situ preparation and immediate loading may have an influence on the clinical outcome.