Abstract
A direct plasma injection HPLC method has been developed for the determination of selected benzodiazepines (nitrazepam, clobazam, oxazepam, lorazepam). The method uses an analytical hydrophobic shielded phase (Hisep) column equipped with a Hisep guard column, are easy to perform and requires 20 ul of a filtered plasma sample. The chromatographic run time is less than 15 min using a mobile phase of 15:85 v/v acetonitrile–0.18 M ammonium acetate pH 2.5. The method is good for 175 injections before replacement of the guard column. The method was linear in the range 0.5–18 ug ml −1 (r>0.99, n=6) for the analytes with R.S.D. less than 10.82%. Interday and intraday variability were found to be less than 14%. The limits of detection and quantitation were 0.16 (s/n>3) and 0.5 ug ml −1 (s/n>10), respectively, for each of the four benzodiazepines.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.