Abstract

Cherenkov emission (CE) is a visible blueish light emitted in water mediums irradiated by most radiotherapy treatment beams. However, CE is produced anisotropically which currently imposes a geometrical constraint uncertainty for dose measurements. In this work, polarization imaging is proposed and described as a method enabling precise 2D dose measurements using CE. CE produced in a water tank is imaged from four polarization angles using a camera coupled to a rotating polarizer. Using Malus’ law, the polarized component of CE is isolated and corrected with Monte Carlo calculated CE polar and azimuthal angular distributions. Projected dose measurements resulting from polarization-corrected CE are compared to equivalent radiochromic film measurements. Overall, agreement between polarized corrected CE signal and films measurements is found to be within 3%, for projected percent depth dose (PPDD) and profiles at the different tested energies (gamma: 6 and 18,hbox {MV}, e^{-}: 6 and 18,hbox {MeV}). In comparison, raw Cherenkov emission presented deviations up 60% for electron beam PPDDs and 20% for photon beams PPDDs. Finally, a degree of linear polarization between 29% and 47% was measured for CE in comparison to 0.2pm 0.3% for scintillation. Hence, polarization imaging is found to be a promising and powerful method for improved radio-luminescent dose measurements with possible extensions to signal separation.