Abstract
Chiral high-performance liquid chromatography was employed for determination of the enantiomeric purity of levodopa and methyldopa. The determination of d-DOPA in levodopa was accomplished using a chiral ligand-exchange chromatograpy with an ordinary C 18 column and a chiral mobile phase containing N, N-dimethyl- l-phenylalanine and Cu(II) acetate or by means of LC on a teicoplanin column in conjunction with ethanol–water (65:35, v/v). Both methods gave good performance, however, the latter was faster and more convenient and suitable for routine analyses. For the determination of d-methyldopa a LC method based on the use of a teicoplanin column in polar organic mode with methanol-acetic acid-triethylamine (1000:0.05:0.05, v/v/v) mobile phase was developed. The precision, accuracy, linearity and selectivity were satisfactory. In comparison with pharmacopoeial polarimetric methods (according to the European Pharmacopoeia and the Pharmacopoea Bohemoslovaca), the LC methods proved to be much more sensitive giving detection limits 0.04% of d-DOPA and 0.3% of d-methyldopa.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.